• ABOUT
    •   Home
    •   Project
    •   Specifications
    •   Team
    •   Reference
  • OBSERVATION
    •   Operation
    •   Preparation
    •   ETC
  • ANALYSIS
    •   Data Analysis
  • LANG 
    •   English
    •   日本語

Basic Instrument Parameters

Wavelength coverage
516 -- 593 nm (fixed)
Spectral resolution
R ∼ 65,000 (see below for more details)
Fiber diameter
130 μm = 2.2 arcsec
Fiber length
35m (from Nasmyth to Spectrograph)
Wavelength calibrators
ThAr lamp, iodine(I2)-cell
Throughput
∼ 2.5-3.0 % (see below for more details)
Detector
one 2k x 4k CCD (e2V CCD44-82BI), 15μm×15μm pixel
Readout system
M-front 2 + Messia-V
Conversion factors
1.92 & 1.95 e-/ADU
Readout noise
∼ 4 e-
System Dark current
∼ 10 e- /pix /hour
Saturation level
> 51,000 ADU (within 0.05% linearity error) (see below for more details)
Total Dead Time for Readout
∼ 30 sec / exposure
Minimum exposure time
1 sec
Maximum exposure time
3,600 sec exposures have been confirmed to work normally.

Optical Layout

Nasmyth Unit (on Seimei Telescope)

Nasmyth Unit
Nasmyth Unit Optical Layout

The entrance to the fiber and the guide camera system are located in a unit mounted at the Nasmyth focus (red) of the telescope. The unit also contains a calibration light source (ThAr, flat) and an I2 cell, which can be inserted into the optical path as required.

Light is introduced into the Nasmyth unit by fixing the rotator angle of the telescope (offset angle -74.9°). Therefore, during guiding, image rotation always occurs depending on the position of the telescope.

The light that entered the fiber from the pinhole is guided into the spectrograph room on the 2nd floor of the dome through a 35m long fiber.

Spectrograph (in the 2nd Floor of the dome)

GAOES-RV Spectrograph

The light (F3) that has passed through the fiber is converted to F30 once and enters into the image slicer. Here, it is divided into five slice images, converted again to F12.2 for the GAOES spectrograph.

This spectrograph room is always kept at 20.0 degC by precision air conditioning, enabling stable wavelength measurement.

GAOES-RV spectrograph Layout

Image Slicer

The same Bowen-Wolraven type slicer element used in the Subaru/HDS and Okayama 188 cm telescope HIDES is used (for details, see Tajitsu et al. 2012, PASJ, 64, 77).

By attaching a plane-parallel plate to the output side of the slicer, five focused slice images can be obtained, and a high dispersion spectrum can be obtained using more incident photons on the fiber.

GAOES-RV Image Slicer

Guide Camera & Target Acquisition

Non-linearity before correction
The target on the fiber is guided by the leaking light of the target itself placed on the fiber.
Please note that if there is a bright object within about 10 arcsec of the target object, it will interfere with the guiding.

The centers of the FoV of the FoV confirmation camera (ZWO Cam) and the guide camera of GAOES-RV are well adjusted to match each other.


FoV
50” × 50” (512 pixel × 512 pixel)
Entrance aperture for the fiber
φ2.2” (130μm, F3)
Filters
ND1 / ND2 / ND3 / 2 × Band-pass filters
Limiting magnitude for Guiding
V < 13 mag with 5 sec exposure

Throughput

Throughput of GAOES-RV spectrograph
The efficiency of GAOES-RV with the telescope is 2.5-3% in the observation wavelength range.
The actual observation efficiency depends on the atmospheric extinction and the efficiency of the fiber inlet,

1 - exp(ln 2 × 2.22 / seeing2)

, will be taken into consideration.
The S/N ratios of objects you want to observe can be estimated with ETC.

CCD Linearity

Non-linearity before correction
We performed the same test as Subaru/HDS by taking sets of flat light frames with two the integration times (13 and 26 sec) and comparing their counts. We have verified that the error from linearity is within 50 e- (0.05 %) at 100,000 e-.

See below for the details of this test.

Tajitsu et al. 2010, "Nonlinearity in the Detector used in the Subaru Telescope High Dispersion Spectrograph", Publ. of National Astronomical Observatory of Japan, 13, 1

Spectral resolution

Reolution scattering on the CCD chip
This is the spectral resolution measured with the spectrum of the ThAr light source obtained after the Hartmann test (measured on February 28, 2023).

Stability of Spectrograph

  • Dispersion direction: about 0.1 pixel (half night to overnight), about 0.5 pixel (different observation periods)
  • Spatial direction: within 0.2 pixels
  • Spectrograph room temperature: 21±0.1 degC

Accuracy of RV measurement

Radial velocity standard
The left figure is the result of applying the standard analysis method (Sato+ 2002) to the observation data using the iodose cell. At present (March 2023), we have achieved an accuracy and stability of about 3m/s for about 2 months for bright solar-type stars.
In the future, by optimizing the analysis method for GAOES-RV and improving the stability of the spectrograph, we aim to achieve the target accuracy (1m/s or less).

In addition to general analysis of high-dispersion spectra, special analysis is required for precise radial velocity measurements using the I2 cell. If you expect to use the analysis code used by the development group, we strongly recommend that you contact the instrument PI and conduct joint research.
RV observations of an exoplanet star

contact to: Akito Tajitsu (Okayama Branch, Subaru Telescope, NAOJ, NINS)      Last modified: Thu Sep 19 13:21:14 JST 2024